Florida Department of Agriculture and Consumer Services

09/15/2025 | Press release | Distributed by Public on 09/15/2025 12:24

Recall - BD Alaris (BD Alaris Pump Module model 8100 used with a subset of compatible pump infusion sets) produced by BD - Pump performance variations compared to the[...]

Summary

Company Announcement Date: September 12, 2025 FDA Publish Date: September 15, 2025 Product Type: Medical Devices Reason for Announcement:
Recall Reason Description
Pump performance variations compared to the performance described in the user manual could impact infusion delivery
Company Name: BD Brand Name:
Brand Name(s)
BD Alaris
Product Description:
Product Description
BD Alaris Pump Module model 8100 used with a subset of compatible pump infusion sets

Company Announcement

FRANKLIN LAKES, N.J. (Sept. 12, 2025) - BD (Becton, Dickinson and Company) (NYSE: BDX), a leading global medical technology company, today expanded their Class I voluntary recall initiated in the United States on July 8, 2025 and expanded on July 17, 2025 to inform customers of worst-case performance, under certain use cases, for the BD Alaris™ Pump Module model 8100 when used with a subset of compatible pump infusion sets, as compared to the performance described in the user manual. The update includes an expansion of 15 additional impacted pump infusion sets included in the recall, as well as new corrective actions and risk information. The 15 additional sets had been previously discontinued but may remain in inventory based on the expiration date.

This issue was identified through internal testing and, to date, BD has not received any complaints associated with this issue. The affected products have the potential to result in patient death or serious adverse events, with a particular higher risk to vulnerable patient populations such as neonates and critically ill patients. The recall has been designated as a Class I recall by the U.S. Food and Drug Administration (FDA).

Pump performance variations as compared to the performance described in the user manual could impact infusion delivery in the following ways:

  • Flow rate and loading bolus dose accuracy (over/under infusion),
  • Upstream and downstream occlusion alarm delay,
  • and post-occlusion bolus volume (POBV) over infusion,

which could adversely impact infusion performance and dose accuracy. The severity and nature of these outcomes depend on the type of medication, fluid, or infusate being administered, as well as the individual patient's condition.

The deviations in performance from the previously published ranges are attributable to these pump infusion sets' design features, such as filters and other in-line components. The products included in the recall were distributed nationwide in the United States, including Guam and Puerto Rico, and in Canada, Belgium and South Africa.

In an update sent to all customers on September 11, 2025, BD shared updated corrective actions, including risk mitigations and recommendations for alternate BD Alaris™ Pump infusion sets to use when available and clinically appropriate. The customer update also included a list of the 15 additional impacted pump infusion sets included in the recall. Information about this recall is available on BD's website or by calling BD at 1-888-562-6018.

Customers should report any complaints experienced with the use of this product to the BD Complaint Center at 1-844-823-5433, Monday - Friday between the hours of 9 a.m. and 6 p.m. ET or by emailing [email protected].

FDA MedWatch Reporting

Adverse reactions/events experienced with the use of this product should be reported to the FDA's MedWatch Program by:

About BD

BD is one of the largest global medical technology companies in the world and is advancing the world of health by improving medical discovery, diagnostics and the delivery of care. The company supports the heroes on the frontlines of health care by developing innovative technology, services and solutions that help advance both clinical therapy for patients and clinical process for health care providers. BD and its more than 70,000 employees have a passion and commitment to help enhance the safety and efficiency of clinicians' care delivery process, enable laboratory scientists to accurately detect disease and advance researchers' capabilities to develop the next generation of diagnostics and therapeutics. BD has a presence in virtually every country and partners with organizations around the world to address some of the most challenging global health issues. By working in close collaboration with customers, BD can help enhance outcomes, lower costs, increase efficiencies, improve safety and expand access to health care. For more information on BD, please visit bd.com or connect with us on LinkedIn at https://www.linkedin.com/company/bd1/, X (formerly Twitter) @BDandCo or Instagram @becton_dickinson.

Contacts:

Media:Fallon McLoughlinDirector, Public [email protected]

Investors:Adam ReiffeSr. Director, Investor [email protected]

Company Contact Information

Consumers: BD Complaint Center 1-844-823-5433
Media: Fallon McLoughlin, Director, Public Relations 201.258.0361 [email protected]
Florida Department of Agriculture and Consumer Services published this content on September 15, 2025, and is solely responsible for the information contained herein. Distributed via Public Technologies (PUBT), unedited and unaltered, on September 15, 2025 at 18:24 UTC. If you believe the information included in the content is inaccurate or outdated and requires editing or removal, please contact us at [email protected]