COMPANY ANNOUNCEMENT
When a company announces a recall, market withdrawal, or safety alert, the FDA posts the company's announcement as a public service. FDA does not endorse either the product or the company.
Read AnnouncementVoluntary Recall - Targeted Pressure Data Reporting Correction - Originally Communicated October 8, 2026
Update Published: November 17, 2026
React Health (3B Medical, Inc.) is providing notice of a voluntary medical device correction (recall) involving specific Luna G3 BPAP System (Model LG3800-30VT) devices due to a data handling issue that may result in incorrect therapy pressure statistical data when the device is operated with Targeted VT enabled.
This notice provides important information for durable medical equipment (DME) providers, hospitals, healthcare facilities, clinicians, and patients using affected devices.
Affected Product Identification and Distribution
| Product name | Luna® G3 BPAP System |
| Model | LG3800-30VT |
| UDI-DI(s) | 06948538365731 |
| Affected serial number(s) | APPENDIX A |
|
Number of affected devices |
7,533 devices distributed |
|
Distribution locations / recipients |
US and Puerto Rico |
Explanation of the Issue
React Health is recalling specific units of the Luna G3 BPAP System (Model LG3800-30VT) due to a data storage and transmission error while Targeted VT is enabled. Devices operating with Targeted VT enabled will not record the actual pressure statistical data delivered to the patient on the internal SD card or transmit the pressure to the cloud-based management system (React Health Connect).
Users that have not enabled the Targeted VT feature on their Model LG3800-30VT device will not experience this issue.
Potential Risk and Clinical Considerations
The hazardous condition requires the device to be operated with Targeted VT enabled and monitored remotely via either data downloaded to the SD card or via wireless communication to the cloud-based management system (React Health Connect). The consequence of the inaccurate pressure data in historical therapy reports could lead a clinician to adjust device settings based on inaccurate therapy pressure data, which could result in delayed therapy until the setting is corrected. Design features that reduce the risk of this issue include the device adjusting pressure delivered to the patient within the clinician defined minimum and maximum pressures delivering a tidal volume equal to the clinician set tidal volume, within tolerance.
To date, React Health has received fifteen (15) complaints with two (2) serious adverse reports associated with this Luna G3 BPAP System (Model LG3800-30VT) Targeted VT enabled issue.
Clinicians should be aware that inaccurate historical pressure data may lead to misinterpretation of therapy performance, adjustments to device settings based on inaccurate information, or delays in recognizing opportunities to optimize patient therapy. Clinicians should consider these limitations when reviewing historical therapy reports and determining whether additional patient assessment or therapy adjustments are appropriate.
Actions for Patients/Users
Patients using a Luna G3 BPAP System (Model LG3800-30VT) Targeted VT enabled feature should do the following:
Actions for Healthcare Professionals and DME Providers
React Health requests that DME providers, hospitals, healthcare facilities, and physicians/clinicians review this information and follow the corrective actions outlined in the original customer recall notification dated October 8, 2026, including as below:
Firmware Correction and Customer Support
React Health has developed firmware version G3-2.00.80 to correct the identified pressure-data reporting issue. Firmware update SD cards are readily available through React Health, allowing affected devices to be updated without requiring permanent device replacement.
React Health is working directly with DME providers and healthcare facilities to coordinate distribution and implementation of the firmware correction. Upon receipt of the completed Customer Response Form, React Health will arrange shipment of the firmware update along with installation instructions and an updated confirmation letter.
This firmware update will allow the collection and display of reported pressure data of delivered therapy while Targeted VT is enabled. All historic data collected with Targeted VT enabled prior to this upgrade will not be changed and will reflect previously recorded non-fluctuating therapy pressures in device storage and the related cloud platform applications.
Affected customers should follow the firmware installation instructions, historical data management procedures, and update confirmation requirements provided in the original recall notification. Following installation of firmware version G3-2.00.80, the original recall instructions direct customers to erase historical usage data through the device's Clinical Menu - Maintenance - Erase Data function.
Additional Information
React Health has communicated recall instructions to affected customers and is coordinating the firmware correction to address the identified issue.
DME providers and healthcare facilities requiring assistance with affected devices, firmware updates, or recall instructions should contact React Health Customer Support at 863-226-6285, Monday through Friday, 8:00 a.m. to 5:00 p.m. Eastern Time.
If an adverse reaction or quality problem has been experienced with the use of this product, it may be reported using one of the following options:
Patients experiencing adverse reactions or quality problems with the use of this device also should report the issues directly to their DME. Alternatively, patients may report to React Health using the Customer Contact Form.
React Health appreciates the cooperation of our healthcare partners in implementing this correction. Patient safety remains our highest priority, and we are committed to supporting our customers and healthcare professionals throughout the recall process.
For media inquiries or additional information regarding this medical device correction, please contact: